Tandem Mass Spectrometry (MS/MS)
Definition
Selecting one ion, fragmenting it, and measuring the fragments. The fragment pattern is far more specific than mass alone, which is why certificates quote transitions rather than a single mass for confirmation.
In more detail
Tandem mass spectrometry weighs a molecule twice. The first stage selects ions of one mass-to-charge ratio, the precursor, and discards everything else. Those selected ions are then broken apart, usually by colliding them with an inert gas, and a second stage measures the masses of the fragments.
On a triple-quadrupole instrument the stages are physical: the first quadrupole (Q1) selects the precursor, a collision cell fragments it, and the third quadrupole (Q3) passes one chosen fragment. A precursor and product pair is called a transition, and monitoring a few transitions at once is multiple reaction monitoring (MRM). A compound has to have the right mass and also break apart in the right way to give a signal, which is far more specific than a single mass.
A peptide fragments mainly along its backbone, at the peptide bonds, giving two families of fragments named b ions (carrying the N-terminus) and y ions (carrying the C-terminus). Because each family steps down the chain one residue at a time, the gaps between fragment masses spell out the sequence, which is how MS/MS can distinguish two peptides of identical overall mass but different residue order.
On a certificate
One or more transitions written precursor › product in m/z, for example 710.4 › 70.1, with the ionisation mode, and a statement that the transition matched a reference standard at the expected retention time.
What this means for research sourcing
Purity and identity verification are central to research reproducibility. Every Volta compound is released against a >99% HPLC purity specification, and a certificate for the batch you receive is available on request.
Related Terms in Quality
Electrospray Ionisation (ESI)
The soft ionisation technique that gets peptides into a mass spectrometer intact, by spraying the solution through a charged needle. It tends to produce several charge states of one molecule rather than a single ion.
Mass-to-Charge Ratio (m/z)
What a mass spectrometer actually measures: mass divided by charge, not mass. A 4,000 Da peptide carrying three charges appears near m/z 1,334, which is why a raw spectrum rarely shows the molecular weight directly.
Charge State
How many charges an ion carries in the mass spectrometer. Peptides commonly appear in several at once, and the series of peaks they produce is what allows the true molecular mass to be calculated.
Deconvolution
The arithmetic that turns a series of multiply-charged peaks back into one molecular mass. It is the step between what the instrument recorded and the single number printed on a certificate.
Adduct
An ion formed by the analyte plus something else it picked up, commonly sodium or potassium. An unrecognised adduct reads as a mass that does not match the expected one, and as an impurity peak that is not an impurity.
Parts Per Million (ppm)
One milligram per kilogram, or one microgram per gram. Residual solvents and elemental impurities are reported this way, and mass spectrometers also quote mass accuracy in ppm, which is a different use of the same unit.
As-Received Basis
A result calculated on the material exactly as it arrived, water and counterion included. It is the honest basis for what is in the vial and reads lower than the same measurement on a dry basis.
Anhydrous Basis
A result recalculated as though the water content had been removed, using a separately measured water figure. Two certificates can disagree by several percent purely because one reports anhydrous and the other as received.