BPC-157 + TB-500 10mg (5+5)Certificate of Analysis VOLT3211801
- COA by CoreMass Laboratories
- Lot VPWV10100
- Report VOLT3211801
- Reported November 13, 2026
- Batch Purity
- 99.70%
- Test method
- HPLC-UV, LC-MS
- Net peptide content
- 11.53 mg
- Identity
- Confirmed

Product
BPC-157 + TB-500 10mg (5+5)
Specifications, chemistry and storage for the vial this certificate tested.
01
Sample Identification
- Laboratory
- CoreMass Laboratories
- Document type
- Certificate of Analysis
- Report number
- VOLT3211801
- Product
- BPC-157 + TB-500 10mg (5+5)
- Lot / batch
- VPWV10100
- Reported date
- November 13, 2026
- Label claim
- 10 mg
- Method
- HPLC-UV, LC-MS
02
Purity & Quality Analysis
Purity (HPLC)
- Batch Purity
- 99.70%
Net peptide content
- Reported
- 11.53 mg net peptide (BPC-157 5.71 mg + TB-500 5.82 mg)
- Share of label claim
- 115.3% of 10 mg
Identity
- Identity
- Confirmed
- Method
- LC-MS
Safety & contaminant panels
- Endotoxins (USP <85> LAL)
- ≤ 0.05 EU/mL
- Microbial contamination (PCR)
- Not detected
- Heavy metals (ICP-MS)
- Below limits
- Fentanyl screen (Targeted screen)
- Not detected
Laboratory notes
Identity confirmed as BPC-157 and Thymosin β4 by LC-MS. Purity is the combined BPC-157 and TB-500 peak area as a share of total integrated peak area by reversed-phase HPLC with UV detection, 99.70% and 99.72% across two vials. Net peptide content was measured for each peptide: BPC-157 5.71 mg and 5.59 mg, TB-500 5.82 mg and 5.62 mg, against a 5 mg + 5 mg label claim. Results relate only to the sample as received.
03
Testing Methods & COA
What this report tested
- Purity, HPLC-UV
- Identity, LC-MS
- Net peptide content, Quantified
- Endotoxins, USP <85> LAL
- Microbial contamination, PCR
- Heavy metals, ICP-MS
- Fentanyl screen, Targeted screen
Batch Purity
99.70%
>99% is the specification every batch is released to. The figure on a certificate is what a laboratory measured for one lot.
Third-party verification
Tested by CoreMass Laboratories, an independent laboratory, and reported on November 13, 2026. Quote the report number to the laboratory, or open its own record of this report, to check the figures against the source rather than against this page.
- Report number
- VOLT3211801
- Reported
- November 13, 2026
- Laboratory
- CoreMass Laboratories
Report VOLT3211801 · Lot VPWV10100
This certificate describes the sample the laboratory received and the lot it names, and no other batch. It is published for laboratory research reference. All materials are supplied for in-vitro research use only and are not for human or animal use.
All certificates of analysis·How to read a certificate of analysis